• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION 500 ML TPN BAG; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION 500 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938738
Device Problems Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
Patient Problem No Patient Involvement (2645)
Event Date 06/23/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a 500ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag had a small split resulting in a leak.This issue was identified during setup and preparation prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H4: the lot was manufactured from april 25, 2019 - april 26, 2019.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection of the photograph showed liquid and droplets apparent inside the pouch.The alleged small slit was not visible in the photograph.The reported condition was verified.The cause of the condition was not determined.Due to the nature of the returned sample.No additional test could be performed.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
500 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10356773
MDR Text Key201394947
Report Number1416980-2020-04631
Device Sequence Number1
Product Code LHI
Combination Product (y/n)Y
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/01/2022
Device Catalogue NumberH938738
Device Lot Number60181161
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/10/2020
Initial Date FDA Received08/03/2020
Supplement Dates Manufacturer Received08/12/2020
Supplement Dates FDA Received09/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-