• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORAX MEDICAL, INC. 1.5T LINX, 15B; ANTI-REFLUX IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORAX MEDICAL, INC. 1.5T LINX, 15B; ANTI-REFLUX IMPLANT Back to Search Results
Catalog Number LXMC15
Device Problem Patient Device Interaction Problem (4001)
Patient Problems Dysphagia/ Odynophagia (1815); Pyrosis/Heartburn (1883); No Code Available (3191)
Event Date 07/02/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).Date sent: 08/03/2020.The dhr for lot 15608 was reviewed.No ncs, defects, or reworks related to the product complaint were found.Additional information was requested, and the following was obtained: on what date did the implant take place? (b)(6) 2017.Lot # of the linx? 15608.Does the patient have any of the allergies to metals? if so, what test have been done to test for metal allergies.No metal allergy.Is the patient currently taking currently taking steroids / immunization drugs? no.Did the patient have any pre-existing dysphagia or other conditions (other than gerd)? no.Was there any hiatal or crural repair done at the same time as the implant? no ¿ nissen takedown with dissection and partial gastrectomy were completed with linx implant.Was mesh used at time of implant? no.What was the reason for removal of the linx device? dysphagia and sensation of heartburn.At the time of removal, was the device found in the correct position/geometry at the time of removal? yes have the symptoms resolved since the device was explanted? yes ¿ although this is early.
 
Event Description
It was reported that a patient had a linx explant performed on (b)(6) 2020.
 
Manufacturer Narrative
(b)(4).Date sent: 08/26/2020.Additional information received: the product code is lxmc15.A suture was placed around the linx device to aid with extraction as it was extracted through one of the port sites.Fifteen beads were again counted once the device was outside of the body consistent with the size that was placed at her original operation.Device analysis: a knotted suture on a wire adjacent to one of the clasp beads was observed during the visual assessment.Per additional information received, the suture was placed during the explant procedure to aid with the explant of the device.Overall review of the device function and dimensions show no anomalies from a device that has been reasonably changed as part of the explant procedure.Visual analysis was consistent with an explanted device, and link length and tensile force were found to meet the applicable specifications.Overall, no analysis conclusions relevant to the patient experience were found.The dhr for lot 15608 was reviewed.No ncs, defects, or reworks related to the product complaint were found.Mfg date: 30/jun/2017; exp date: 30/jun/2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.5T LINX, 15B
Type of Device
ANTI-REFLUX IMPLANT
Manufacturer (Section D)
TORAX MEDICAL, INC.
4188 lexington avenue north
shoreview MN
MDR Report Key10356776
MDR Text Key201690134
Report Number3008766073-2020-00112
Device Sequence Number1
Product Code LEI
Combination Product (y/n)N
PMA/PMN Number
P100049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 07/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/30/2021
Device Catalogue NumberLXMC15
Device Lot Number15608
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/11/2020
Initial Date Manufacturer Received 07/13/2020
Initial Date FDA Received08/03/2020
Supplement Dates Manufacturer Received08/11/2020
Supplement Dates FDA Received08/26/2020
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-