Model Number AAS00161-22 |
Device Problems
Display or Visual Feedback Problem (1184); Application Program Problem (2880)
|
Patient Problem
No Code Available (3191)
|
Event Date 07/16/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
Out of an abundance of caution, superdimension is filing this mdr due to the additional risk associated with multiple exposures to general anesthesia.(b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during the procedure, the system was set up, the bronchoscope video worked well but when the registration started, the video disappeared.The procedure was cancelled.The patient was under general anesthesia.
|
|
Manufacturer Narrative
|
The incident device has been received and is under evaluation.When the device evaluation is complete a follow-up report will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during the procedure, the system was set up, the bronchoscope video worked well but when the registration started, the video disappeared.The procedure was cancelled and was not rescheduled.The patient was under general anesthesia.
|
|
Manufacturer Narrative
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the unit had data failure.A screen capture file was corrupt.The issue was resolved by installing a system computer hard disk image (re-imaging).It was reported that there were bronch imaging issues with the system monitor.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.A procedure recordings and the device were available for e valuation.Visual inspection noted that the unit had connectivity issues.The issue was resolved by installing a system computer hard disk image and reconnecting the bnc video connection to the superdimension system.It was reported that the procedure was cancelled due to a system issue.The reported issue could not be confirmed.The most likely cause could not be established from the information available.It was also reported that there were bronch imaging issues with the system monitor.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur if video server was not available.If the video server was unavailable, it can potentially affect the display of video input during the procedure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|