Model Number 2545-00-060
Device Problems Device Slipped (1584); Device-Device Incompatibility (2919); Unintended Movement (3026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/28/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Attune pin puller in loan set not latching onto attune pinning system.Surgeon able to still pull pins out throughout case with tool but reported handle was slightly loose.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: after examination of the returned photos the failure mode could not be confirmed.Root cause undetermined.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PIN PULLER
Type of Device
ATTUNE INSTRUMENTS : EXTRACTION INSTRUMENTS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key10360003
MDR Text Key201550069
Report Number1818910-2020-17413
Device Sequence Number1
Product Code HWB
UDI-Device Identifier10603295130574
UDI-Public10603295130574
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2545-00-060
Device Catalogue Number254500060
Device Lot NumberPG246438
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/28/2020
Initial Date FDA Received08/03/2020
Supplement Dates Manufacturer Received10/05/2020
Supplement Dates FDA Received10/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-