• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII MIS DCF DISTAL CUT BLK; PROSTHESIS,KNEE,FEMOROTIBIAL,SEMI-CONSTRAINED,CEMENTED,METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII MIS DCF DISTAL CUT BLK; PROSTHESIS,KNEE,FEMOROTIBIAL,SEMI-CONSTRAINED,CEMENTED,METAL/POLYMER Back to Search Results
Model Number 71441147
Device Problem Failure to Disconnect (2541)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/09/2020
Event Type  malfunction  
Event Description
It was reported that the pin got stuck while attempting to introduce it into the cutting block and was unable to remove it.No delay reported.No patient injuries reported.An s&n backup was available.
 
Manufacturer Narrative
The associated device, intended for use in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.A review of complaint history on the listed part revealed, no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal, a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend, that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation warranted for this complaint.However, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII MIS DCF DISTAL CUT BLK
Type of Device
PROSTHESIS,KNEE,FEMOROTIBIAL,SEMI-CONSTRAINED,CEMENTED,METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10360171
MDR Text Key201496071
Report Number1020279-2020-03670
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010497185
UDI-Public03596010497185
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 05/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71441147
Device Catalogue Number71441147
Device Lot Number05AM01028
Initial Date Manufacturer Received 07/09/2020
Initial Date FDA Received08/03/2020
Supplement Dates Manufacturer Received05/04/2021
Supplement Dates FDA Received05/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-