Model Number MAJ-1516 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/10/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The subject device has not returned to olympus medical systems corp.(omsc) for evaluation.There were no further details provided.If significant additional information is received, this report will be supplemented.
|
|
Event Description
|
During inspection prior to use, the device's lock lever didn't fit in with automated endoscope reprocessor.The customer used another connecting tube to complete the reprocessing.There was no report of patient injury associated with the event.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information.The subject device was returned to olympus medical systems corp.(omsc) for evaluation.Omsc could not review the manufacturing history (dhr) of the subject device because the lot number of the device is unknown.The exact cause of the reported phenomenon could not be conclusively determined.However, based upon the investigation result by omsc there was the possibility that the reported phenomenon was attributed to the accidental external force applied to the lock lever such as during transportation.
|
|