Model Number B-50000
Device Problem Patient-Device Incompatibility (2682)
Patient Problems Nausea (1970); Vomiting (2144); Abdominal Distention (2601)
Event Date 01/31/2020
Event Type  malfunction  
Manufacturer Narrative
The device will not be returned for analysis.A review of the device labeling notes the following: the risk documentation for this device establishes the occurrence ranking for the reported event "inflation" as "remote", which is defined by apollo as.02% -.49% of complaints over units sold.A review for the intragastric balloon products for the failure mode "inflation" is performed during quarterly complaint analysis meetings (cam) and has demonstrated that global balloon inflation rate remains at "remote".Therefore, apollo determined that the reported event is occurring within the range of the expected frequency and severity for this reported event, as referenced in the cam meeting slides.The current orbera® intragastric balloon system directions for use (dfu) addresses the known and anticipated potential events of "vomiting", "bloating" and "nausea" as follows: "the physiological response of the patient to the presence of the orbera system balloon may vary depending upon the patient's general condition and the level and type of activity.The types and frequency of administration of drugs or diet supplements and the overall diet of the patient may also affect the response." "each patient must be monitored closely during the entire term of treatment in order to detect the development of possible complications.Each patient should be instructed regarding symptoms of deflation, gastrointestinal obstruction, ulceration and other complications which might occur, and should be advised to contact his/her physician immediately upon the onset of such symptoms." "complications: possible complications of the use of the orbera system include: gastric discomfort, feelings of nausea and vomiting following balloon placement as the digestive system adjusts to the presence of the balloon.Continuing nausea and vomiting.This could result from direct irritation of the lining of the stomach or as a result of the balloon blocking the outlet of the stomach.It is even theoretically possible that the balloon could prevent vomiting (not nausea or retching) by blocking the inlet to the stomach from the esophagus.Abdominal or back pain, either steady or cyclic.".
 
Event Description
Patient presented nausea, vomiting, abdominal distention.Inflation of orbera balloon.
 
Manufacturer Narrative
Supplement #1 medwatch submitted to the fda on 05/feb/2021.Device evaluation summary: the device was returned to the apollo device analysis laboratory on (b)(6) 2021.Received a deflated balloon with a significant amount of brown discoloration.There are two slits on the shell seen by the naked eye.Under microscopic analysis, the slits on the shell have jagged edges which is consistent with a surgical instrument for removal purposes.Due to the slits on the shell, functional evaluation of the device could not be conducted.The complaint could not be verified as it is uncertain how the inflation occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORBERA INTRAGASTRIC BALLOON SYSTEM
Type of Device
INTRAGASTRIC BALLOON
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC.
1120 s. capital of texas hwy
bldg 1, ste. 300
austin, tx
MDR Report Key10363447
MDR Text Key201603974
Report Number3006722112-2020-00069
Device Sequence Number1
Product Code LTI
Combination Product (y/n)N
PMA/PMN Number
P140008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 04/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/03/2021
Device Model NumberB-50000
Device Catalogue NumberB-50000
Device Lot NumberAF03047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2021
Initial Date Manufacturer Received 04/22/2020
Initial Date FDA Received08/04/2020
Supplement Dates Manufacturer Received01/07/2021
Supplement Dates FDA Received02/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ESOMEPRAZOLE 40 MG; ESOMEPRAZOLE 40 MG
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age49 YR
Patient Weight86
-
-