• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII MIS DCF DISTAL CUT BLK; PRSTHSS,KNEE,PTLL/FMRTBL,SM-CNSTRND,UNCMNTD,PRS,CTD,PLYMR/MTL/PLYMR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII MIS DCF DISTAL CUT BLK; PRSTHSS,KNEE,PTLL/FMRTBL,SM-CNSTRND,UNCMNTD,PRS,CTD,PLYMR/MTL/PLYMR Back to Search Results
Catalog Number 71441147
Device Problem Mechanical Jam (2983)
Patient Problem No Patient Involvement (2645)
Event Date 07/09/2020
Event Type  malfunction  
Event Description
It was reported that during instrument inspection was found that the pin is stuck in the hole of the genesis ii mis dcf distal cutting block.There was no case involved.
 
Manufacturer Narrative
The device, intended use in treatment, was returned for evaluation.A visual inspection confirmed a pin is stuck in the gii mis dcf distal cut blk.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A review of complaint history for the listed part revealed no prior complaints for the listed batch.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII MIS DCF DISTAL CUT BLK
Type of Device
PRSTHSS,KNEE,PTLL/FMRTBL,SM-CNSTRND,UNCMNTD,PRS,CTD,PLYMR/MTL/PLYMR
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10363615
MDR Text Key201589293
Report Number1020279-2020-03693
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 11/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71441147
Device Lot Number13GM16541
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 07/09/2020
Initial Date FDA Received08/04/2020
Supplement Dates Manufacturer Received11/03/2020
Supplement Dates FDA Received11/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-