• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALERE SAN DIEGO, INC. CONSULT HCG URIN/SRM COMBO 5002 25; PREGNANCY TEST

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALERE SAN DIEGO, INC. CONSULT HCG URIN/SRM COMBO 5002 25; PREGNANCY TEST Back to Search Results
Model Number FHC-A202
Device Problem False Negative Result (1225)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Customer / distributor unwilling to return / unresponsive.Investigation conclusion: an investigation was performed on retention devices from the reported lot number.Retention devices were tested with qc cut-off urine standard (20 miu/ml), qc cut-off serum standard (10 miu/ml), and high hcg (215.2 - 232.6 iu/ml) clinical urine.Results were read at 3 minutes for urine samples and at 5 minutes for serum samples.All devices produced expected positive results.No false negative results were observed.The case details were reviewed along with the complaint history for the reported issue and no indications of a systemic issue were identified.Batch record review found no relevant non-conformances; the lot met all final release specifications.Review of the risk management report for this product found that the reported issue is within the risk profile for this device; no new hazard has been identified.The reported complaint was not replicated.Limited details regarding patient, product, and testing information could be obtained.A probable cause could not be identified based on the information available.Per the package insert: very dilute urine specimens, as indicated by a low specific gravity, may not contain representative levels of hcg.False negative results may occur when the levels of hcg are below the sensitivity level of the test.When pregnancy is still suspected, a first morning urine specimen should be collected 48 hours later and tested.This test provides a presumptive diagnosis for pregnancy.A confirmed pregnancy diagnosis should only be made by a physician after all clinical and laboratory findings have been evaluated.
 
Event Description
Unspecified date: false negative result on the consult diagnostics hcg urine/serum combo cassette kit.No adverse patient outcomes reported.Although requested, no further information was able to be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSULT HCG URIN/SRM COMBO 5002 25
Type of Device
PREGNANCY TEST
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
9975 summers ridge road
san diego, ca
Manufacturer Contact
monika burrell
9975 summers ridge road
san diego, ca 
8052506
MDR Report Key10370900
MDR Text Key201871389
Report Number2027969-2020-00042
Device Sequence Number1
Product Code JHI
UDI-Device Identifier20612479202768
UDI-Public(01)20612479202768(17)210831(10)HCG9082044
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2021
Device Model NumberFHC-A202
Device Lot NumberHCG9082044
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/27/2020
Initial Date FDA Received08/05/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-