SMITH & NEPHEW, INC. FOOTPRINT ULTRA PK SUTURE ANCHOR 4.5; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72202901
Device Problem Misfire (2532)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/13/2020
Event Type  malfunction  
Event Description
It was reported that during a shoulder surgery using a footprint device, the third anchor was difficult to pass it through the eyelet, after a few attempts it passed but when it was trying to be inserted into the bone hole, the anchor fell off.The anchor was retrieved and discarded.The procedure was completed without a significant delay using a back-up device in the same bone hole.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3,h6: the reported 4.5 footprint ultra pk suture anchor assembly, used in treatment, will not be returned for evaluation.Without the reported product a visual evaluation cannot be performed and the customers complaint cannot be confirmed.From the information provided, the anchor fell off the inserter during use.An exact root cause cannot be determined with confidence; however, factors that could have contributed to the reported event include: inadvertent dislodgment of the anchor during use.The instruction for use outlines precautionary statements and instructions in regards to the use of the device to avoid damage or non-functionality.A review of the manufacturing and complaint records was performed for the reported lot, there were no indications that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOOTPRINT ULTRA PK SUTURE ANCHOR 4.5
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key10376491
MDR Text Key202002272
Report Number1219602-2020-01185
Device Sequence Number1
Product Code MBI
UDI-Device Identifier03596010650528
UDI-Public03596010650528
Combination Product (y/n)N
PMA/PMN Number
K093897
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72202901
Device Catalogue Number72202901
Device Lot Number2047171
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/13/2020
Initial Date FDA Received08/06/2020
Supplement Dates Manufacturer Received08/28/2020
Supplement Dates FDA Received09/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-