• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. VISERA ELITE XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. VISERA ELITE XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S190
Device Problem Defective Device (2588)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/16/2020
Event Type  malfunction  
Manufacturer Narrative
The subject device has not returned to olympus medical systems corp.(omsc) for evaluation.There were no further details provided.If significant additional information is received, this report will be supplemented.
 
Event Description
During the preparation for an unspecified procedure using the subject device, the front panel display of the subject device blinked.Other detailed information was not provided.There was no report of patient injury associated with the event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Olympus medical systems corp.(omsc) could not investigate the device, because the device was not returned to omsc.Omsc reviewed the manufacturing history (dhr) of the device and confirmed no irregularity.Since the device was not returned, the exact cause was unknown; however, the following was supposed to be the cause.- the power supply unit of the device was broken by some cause.- the light guide connection detection function was failed by some cause.The instruction manual of the subject device states the corresponding method in case of an abnormality.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA ELITE XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10377197
MDR Text Key202080534
Report Number8010047-2020-05360
Device Sequence Number1
Product Code NWB
Combination Product (y/n)N
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 10/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberCLV-S190
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/19/2020
Initial Date FDA Received08/06/2020
Supplement Dates Manufacturer Received09/29/2020
Supplement Dates FDA Received10/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-