This event occurred in (b)(6).The customer's meters and strips were requested for investigation.The product has not been received at this time.Retention samples glucose lot 37996005 were measured with controls gkl 60 lot 346092-99 /gkl 180 lot 346093-99 and 3 edta blood on reference devices compared to laboratory method.Ranges control solutions: range gkl 60: 58 ¿ 98 mg/dl investigation results using retention material: qc 1: 82 mg/dl, qc 2: 74 mg/dl, qc 3: 75 mg/dl.Investigation results using retention material: range gkl 180: 173 ¿ 232 mg/dl qc 1: 200 mg/dl, qc 2: 192 mg/dl, qc 3: 204 mg/dl.All results were within the control range.No significant differences to the laboratory method.Occupation is lay user/patient.The investigation is ongoing.
|