THERAKOS INC. CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Model Number NOT APPLICABLE
Device Problems Break (1069); Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/09/2020
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction drive tube leak/break.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot h168 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot h168 shows no trends.Trends were reviewed for complaint categories, drive tube leak/break and alarm #7: blood leak? (centrifuge chamber).No trends were detected for these complaint categories.The assessment is based on the information available at the time of investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.(b)(4).S.D.A.07-aug-2020.
 
Event Description
The customer contacted mallinckrodt to report that they experienced a drive tube leak/break with their cellex photopheresis kit ("kit") during an extracorporeal photopheresis (ecp) treatment.The customer stated that at the beginning of the treatment they received an alarm #7: blood leak (centrifuge chamber) alarm and a leak on the drive tube was observed.The customer aborted the ecp treatment and did not return residual blood within the kit to the patient.The patient was reported to be in stable condition.The customer discarded the kit and did not return product for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
THERAKOS INC.
bedminster, nj
Manufacturer (Section G)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
college business & tech park
cruiserath road
blanchardstown, dublin, D15 T X2V
EI   D15 TX2V
Manufacturer Contact
megan vernak
1425 us route 206
bedminster, nj 
MDR Report Key10378210
MDR Text Key241305398
Report Number2523595-2020-00090
Device Sequence Number1
Product Code LNR
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 08/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/01/2021
Device Model NumberNOT APPLICABLE
Device Catalogue NumberCLXECP
Device Lot NumberH168
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/09/2020
Initial Date FDA Received08/07/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/02/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-