BARD ACCESS SYSTEMS M.R.I. LOW-PROFILE IMPLANTABLE PORT, HICKMEN SINGLE-LUMEN, 6.6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number 0603880C
Device Problems Loss of or Failure to Bond (1068); Fracture (1260); Migration or Expulsion of Device (1395); Suction Problem (2170)
Patient Problem Injury (2348)
Event Date 04/23/2020
Event Type  Injury  
Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/ patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Expiry date: (08/2021).The catalog number has not been cleared in the us but is similar to the m.R.I.Low-profile implantable port, hickmen single-lumen, 6.6f that are cleared in the us.The pro code and 510k number for the m.R.I.Low-profile implantable port, hickmen single-lumen, 6.6f is identified.
 
Event Description
It was reported that during a palliative chemotherapy treatment, the catheter allegedly migrated to pulmonary artery and dislodged.The catheter was removed through surgery.The patient was transferred to cardiothoracic surgery for further treatment.The patient status was unknown.
 
Manufacturer Narrative
H10: manufacturing review: the device history record review was performed for the reported lot number and this lot meets all release criteria.The manufacturing review did not indicate any possible manufacturing issue that could be related to the reported event.Investigation summary: the device was not returned for evaluation.Therefore, the reported catheter dislodgement, catheter migration, no blood return issue and catheter obstruction cannot be confirmed.A device history record review could not be performed as the lot number is unknown.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 08/2021),g4,h6(device: 2423).H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the m.R.I.Low-profile implantable port, hickmen single-lumen, 6.6f that are cleared in the us.The pro code and 510k number for the m.R.I.Low-profile implantable port, hickmen single-lumen, 6.6f is identified in d2 and g5.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during a palliative chemotherapy treatment, the catheter allegedly migrated to pulmonary artery and dislodged.The catheter was removed through surgery.The patient was transferred to cardiothoracic surgery for further treatment.The patient status was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.R.I. LOW-PROFILE IMPLANTABLE PORT, HICKMEN SINGLE-LUMEN, 6.6F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key10379104
MDR Text Key202027478
Report Number3006260740-2020-02849
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741111167
UDI-Public(01)00801741111167
Combination Product (y/n)N
PMA/PMN Number
K924250
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0603880C
Device Lot NumberRECW0346
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/16/2020
Initial Date FDA Received08/07/2020
Supplement Dates Manufacturer Received11/18/2020
Supplement Dates FDA Received11/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-