BAXTER HEALTHCARE CORPORATION UV FLASH TRANSFER SETS; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number T5C4326 |
Device Problem
Fitting Problem (2183)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/14/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the catheter adaptor (patient connector) of a uv-flash solution trans set (short) - (b)(6) was ¿idled¿ and could not connect to the titanium adapter.The nurse observed this while preparing the patient for peritoneal dialysis therapy when the patient visited the hospital for a periodic transfer set replacement.As a result, the transfer set was replaced.The titanium adapter was not alleged and was still in use.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
Additional information in h3, h6.H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing including leak, clear passage, and clamp function testing was performed with no issues noted.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|