BAXTER HEALTHCARE CORPORATION UV FLASH TRANSFER SETS; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Catalog Number T5C4326
Device Problem Fitting Problem (2183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/14/2020
Event Type  malfunction  
Manufacturer Narrative
The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the catheter adaptor (patient connector) of a uv-flash solution trans set (short) - (b)(6) was ¿idled¿ and could not connect to the titanium adapter.The nurse observed this while preparing the patient for peritoneal dialysis therapy when the patient visited the hospital for a periodic transfer set replacement.As a result, the transfer set was replaced.The titanium adapter was not alleged and was still in use.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information in h3, h6.H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing including leak, clear passage, and clamp function testing was performed with no issues noted.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UV FLASH TRANSFER SETS
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10383008
MDR Text Key202165692
Report Number1416980-2020-04822
Device Sequence Number1
Product Code KDJ
UDI-Device Identifier04987456422852
UDI-Public(01)04987456422852
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/23/2024
Device Catalogue NumberT5C4326
Device Lot NumberH19A23058
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2020
Initial Date Manufacturer Received 07/15/2020
Initial Date FDA Received08/08/2020
Supplement Dates Manufacturer Received09/08/2020
Supplement Dates FDA Received09/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
TITANIUM ADAPTER.
-
-