SMITH & NEPHEW, INC. STEM IMPACTOR ROD; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71364011
Device Problem Degraded (1153)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/14/2020
Event Type  malfunction  
Event Description
It was reported that during a thr the device threads on rod got stripped out.The malfunction happened outside of the patient.There was no delay nor injury reported.The surgery was finished with the same device.
 
Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection confirmed the threads are damaged on the stem impactor rod.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A review of complaint history for the listed part revealed no prior complaints for the listed batch/failure mode.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEM IMPACTOR ROD
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10383187
MDR Text Key202163263
Report Number1020279-2020-03807
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010199270
UDI-Public03596010199270
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 08/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71364011
Device Catalogue Number71364011
Device Lot Number14AM08134
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2020
Initial Date Manufacturer Received 07/14/2020
Initial Date FDA Received08/09/2020
Supplement Dates Manufacturer Received08/04/2020
Supplement Dates FDA Received08/18/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-