Model Number I.V. STATION
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/20/2020
Event Type  malfunction  
Event Description
On (b)(6) 2020 a preparation that was compounded by an i.V.Station device had a label affixed that contained incorrect content.The beyond use date (bud) printed on the label exceeded the expiration date provided by the vial manufacturer.Specifically, norepinephrine drug preparations were labeled with a bud of october 18, 2020 while the drug vial indicated an expiration date of september 30, 2020.The software, which is intended to consider vial expiration dates when calculating the bud, did not perform as intended and applied a date beyond allowable use.The mislabeled preparations were identified and there are no known adverse patient effects.
 
Manufacturer Narrative
As of (b)(6) 2020, the establishment registration and listing for this device was updated to omnicell, inc.From aesynt, inc, which was not reflected in the original report submission.Therefore, this report is a correction to the manufacturer listed in sections d3 and g1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNICELL I.V.STATION
Type of Device
PHARMACY COMPOUNDING DEVICE
Manufacturer (Section D)
OMNICELL, INC.
51 pennwood place
warrendale PA 15086
MDR Report Key10394406
MDR Text Key202588179
Report Number3011278888-2020-00017
Device Sequence Number1
Product Code NEP
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 05/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberI.V. STATION
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 07/20/2020
Initial Date FDA Received08/11/2020
Supplement Dates Manufacturer Received05/07/2021
Supplement Dates FDA Received05/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-