Model Number 6021-0637
Device Problem Unstable (1667)
Patient Problem Injury (2348)
Event Date 07/24/2020
Event Type  Injury  
Manufacturer Narrative
Review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event Description
Patient went to er because he felt a snap and his hip became unstable.Radiographs were taken and his hip, radiographs revealed an issue with head alignment.The patient contacted the (b)(6) clinic for an examination.The doctor saw the patient and immediately scheduled him for a hip revision the following day.We revised the head, stem and the acetabulum poly.Patient had his accolade tmzf stem revised.A new head and a new liner was implanted.The trident cup was not damaged and was deemed stable during the surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE PLUS TMZF HIP STEM #6
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
IE   NA
Manufacturer Contact
alessandra chavez
2555 davie road
fort lauderdale, FL 33317
9546280700
MDR Report Key10400551
MDR Text Key202688810
Report Number0002249697-2020-01656
Device Sequence Number1
Product Code LPH
UDI-Device Identifier04546540510716
UDI-Public04546540510716
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121308
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 08/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2011
Device Model Number6021-0637
Device Catalogue Number6021-0637
Device Lot Number20385501
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/24/2020
Initial Date FDA Received08/13/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/31/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age69 YR
-
-