STRYKER ORTHOPAEDICS-MAHWAH V40 COCR LFIT HEAD 40MM/+8; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 6260-9-340
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Injury (2348)
Event Date 07/24/2020
Event Type  Injury  
Manufacturer Narrative
Review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.Device not returned.
 
Event Description
Patient went to er because he felt a snap and his hip became unstable.Radiographs were taken and his hip, radiographs revealed an issue with head alignment.The patient contacted the hughston clinic for an examination.The doctor saw the patient and immediately scheduled him for a hip revision the following day.We revised the head, stem and the acetabulum poly.Patient had his accolade tmzf stem revised.A new head and a new liner was implanted.The trident cup was not damaged and was deemed stable during the surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V40 COCR LFIT HEAD 40MM/+8
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
alessandra chavez
2555 davie road
fort lauderdale, FL 33317
9546280700
MDR Report Key10400552
MDR Text Key202682236
Report Number0002249697-2020-01657
Device Sequence Number1
Product Code LPH
UDI-Device Identifier07613327032765
UDI-Public07613327032765
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061434
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial
Report Date 08/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/31/2011
Device Model Number6260-9-340
Device Catalogue Number6260-9-340
Device Lot Number818MTA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/24/2020
Initial Date FDA Received08/13/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/27/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number2249697-08/29/2016-007R
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age69 YR
-
-