A d-dimer result was obtained on a patient sample on a sysmex ca-660 system using innovance d-dimer reagent.The customer reran the sample for d-dimer using the d-dimer+ (ddi+) application with the same system and reagent, recovering lower.A second sample from the same patient was run for d-dimer using the same system and reagent, recovering higher.The redrawn sample was repeated again for d-dimer using the ddi+ application with the same system and reagent, recovering lower.The higher d-dimer result was reported to the physician(s).It is unknown which d-dimer result was considered correct.There are no reports of patient intervention or adverse health consequences due to the discordant d-dimer result(s).
|