• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI, INC ROTATING CF RESECTOSCOPE INNER SHEATH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC ROTATING CF RESECTOSCOPE INNER SHEATH Back to Search Results
Model Number ERIS-CF27
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus (b)(4) for evaluation.The customer¿s complaint was confirmed.The sheath was returned with the ceramic tip broken off.The broken portion was not returned.The cause of the reported event could not be determined; however, the most likely root cause could be attributed to user handling.The instruction manual contains several warning statements in an effort to prevent damage to the ceramic beak of the device.¿always keep sheaths parallel to one another when assembling and disassembling.If the inner sheath is inserted or removed at an angle to the outer sheath, the lateral force applied to the inner sheath may crack, loosen, break, or otherwise damage the sheath¿s insulated distal tip.A broken tip, or fragments of a damaged tip, can potentially pass through the outer sheath and into the patient.".
 
Event Description
The service center was informed that the ceramic tip broke off the inner sheath.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTATING CF RESECTOSCOPE INNER SHEATH
Type of Device
ROTATING CF RESECTOSCOPE INNER SHEATH
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough MA 01772
Manufacturer Contact
kenneth pittman
93 north pleasant st.
norwalk, OH 44857
9013785969
MDR Report Key10405337
MDR Text Key218005238
Report Number1519132-2020-00050
Device Sequence Number1
Product Code FAS
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberERIS-CF27
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/20/2020
Initial Date FDA Received08/13/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-