Model Number VTICM5_13.7
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/14/2020
Event Type  malfunction  
Manufacturer Narrative
Pma/510k: this product is not marketed in the us.Work order search: no similar complaint type events were reported for units within the same lot.Claim # (b)(4).
 
Event Description
The reporter indicated that a 13.7mm, vticm5_13.7, -10.5/5.5/102 (sphere/cylinder/axis), implantable collamer lens was implanted into the patients right eye (od) on (b)(6) 2020.The patient experienced excessive vault, lens remains implanted.Additional information has been requested but none has been forthcoming.If additional information is received a supplemental medwatch report will be submitted.
 
Manufacturer Narrative
Additional information: b5: "patient's last visit was on (b)(6) 2020.Bcva 20/20; vault 900.Patient is happy.Lens remains implanted." (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia, ca
MDR Report Key10406783
MDR Text Key202947655
Report Number2023826-2020-01751
Device Sequence Number1
Product Code QCB
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 07/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2020
Device Model NumberVTICM5_13.7
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/17/2020
Initial Date FDA Received08/14/2020
Supplement Dates Manufacturer Received09/08/2020
Supplement Dates FDA Received05/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CARTRIDGE MODEL: SFC-45 W/FTP, LOT# UNK.; INJECTOR MODEL: MSI-PF, LOT# UNK.; CARTRIDGE MODEL: SFC-45 W/FTP, LOT# UNK; INJECTOR MODEL: MSI-PF, LOT# UNK
Patient Age26 YR
-
-