MAUDE Adverse Event Report: SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 102096-67A
Device Problem Failure to Sense (1559)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/16/2020
Event Type  malfunction  
Manufacturer Narrative
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
 
Event Description
On (b)(6) 2020,senseonics was made aware of an incident where user received an early sensor replacement alert resulting in an early sensor removal.
 
Manufacturer Narrative
The sensor was investigated and the issue was confirmed during qc testing in-house.Intermittent sensor led failures were detected during this test.The investigation shows the system correctly disabled the sensor due to performance failure and the system's self-test functions are working normally.D8 was this device serviced by a third party? no.D9 device available for evaluation? yes, device received on 10/15/2020.H3.Device evaluated by manufacturer? yes.H6.Health effect - clinical code updated to 4582.H6.Health effect -impact code updated to 2199.H6.Medical device problem code updated to 1559.H6.Component code updated to 510.H6.Type of investigation updated to 4111 and 10.H6.Investigation findings updated to 4203.H6.Investigation conclusions updated to 4307.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown, md
MDR Report Key10408494
MDR Text Key202993415
Report Number3009862700-2020-00451
Device Sequence Number1
Product Code QCD
UDI-Device Identifier00817491022349
UDI-Public00817491022349
Combination Product (y/n)N
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 07/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/10/2020
Device Model Number102096-67A
Device Catalogue NumberFG-4200-00-301
Device Lot NumberWP07582
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2020
Initial Date Manufacturer Received 07/16/2020
Initial Date FDA Received08/14/2020
Supplement Dates Manufacturer Received07/16/2020
Supplement Dates FDA Received12/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age52 YR
-
-