Model Number 130760142
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Bone Fracture(s) (1870); Injury (2348); No Code Available (3191)
Event Date 08/05/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Patient with delta xtend implants insitu, fractured scapula resulting in loose metaglene and glenosphere.Patient has rheumatoid arthritis and identified by the surgeon to have poor bone quality which he feels contributed to the implant failing.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot number was not provided.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND GLENOSPHERE STD D42MM
Type of Device
DELTA XTEND IMPLANTS : SHOULDER GLENOSPHERE
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key10410963
MDR Text Key203034851
Report Number1818910-2020-18193
Device Sequence Number1
Product Code HSD
UDI-Device Identifier10603295027775
UDI-Public10603295027775
Combination Product (y/n)N
PMA/PMN Number
D165018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number130760142
Device Catalogue Number130760142
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/05/2020
Initial Date FDA Received08/14/2020
Supplement Dates Manufacturer Received10/19/2020
Supplement Dates FDA Received10/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
DXTEND HIGHMO PE CUP D42 +3MM; METAGLENE +10MM; UNKNOWN SHOULDER LOCKING SCREW; UNKNOWN SHOULDER LOCKING SCREW; DXTEND HIGHMO PE CUP D42 +3MM; METAGLENE +10MM; UNKNOWN SHOULDER LOCKING SCREW; UNKNOWN SHOULDER LOCKING SCREW
Patient Outcome(s) Required Intervention;
Patient Age76 YR
-
-