Model Number VTICM5_12.6 |
Device Problem
Inadequacy of Device Shape and/or Size (1583)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/25/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
This product is manufactured in the u.S.But not marketed in the u.S.No similar complaint was reported for units within the same lot.(b)(4).
|
|
Event Description
|
The reporter indicated the surgeon implanted a 12.6mm vticm5_12.6 implantable collamer lens, -14.0/+3.0/088 (sphere/cylinder/axis), in the patients left eye (os) on (b)(6) 2020.The lens had low vaulting and remained implanted.The plan is to explant/exchange the lens at a later date.The cause of the event was unknown.
|
|
Manufacturer Narrative
|
B5: the lens was explanted on (b)(6) 2020 and was exchanged for a longer lens.The lens exchange resolved the problem.Claim # (b)(4).
|
|
Manufacturer Narrative
|
Additional information: h3-device evaluation: lens returned in a micro-centrifuge vial with moisture.Visual inspection found no visible damage to lens and residue on lens.(b)(4).
|
|