OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA III XENON LIGHT SOURCE Back to Search Results
Model Number CLV-190
Device Problem Failure to Shut Off (2939)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The suspect device was returned to olympus and evaluated.The reported problem was confirmed.The evaluation determined the power switch was broken and an upgrade to a new version switch was required.
 
Event Description
Olympus received a report from the user facility that the power button was stuck in the on position.There was no patient injury or harm, associated with the problem, reported to olympus.
 
Manufacturer Narrative
This supplemental report is submitted to provide the results of the legal manufacturer¿s investigation and device history record (dhr) review.The dhr for the subject device was reviewed and it was verified the device was manufactured in accordance with documented specifications.The legal manufacturer performed an investigation.A conclusive root cause was not identified.The legal manufacturer determined the likely cause was damage to the power switch related to the design.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10417505
MDR Text Key223647415
Report Number8010047-2020-05527
Device Sequence Number1
Product Code NWB
Combination Product (y/n)N
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 01/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/21/2020
Initial Date FDA Received08/17/2020
Supplement Dates Manufacturer Received12/07/2020
Supplement Dates FDA Received01/04/2021
Is This a Reprocessed and Reused Single-Use Device? No