Model Number CLV-190 |
Device Problem
Failure to Shut Off (2939)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The suspect device was returned to olympus and evaluated.The reported problem was confirmed.The evaluation determined the power switch was broken and an upgrade to a new version switch was required.
|
|
Event Description
|
Olympus received a report from the user facility that the power button was stuck in the on position.There was no patient injury or harm, associated with the problem, reported to olympus.
|
|
Manufacturer Narrative
|
This supplemental report is submitted to provide the results of the legal manufacturer¿s investigation and device history record (dhr) review.The dhr for the subject device was reviewed and it was verified the device was manufactured in accordance with documented specifications.The legal manufacturer performed an investigation.A conclusive root cause was not identified.The legal manufacturer determined the likely cause was damage to the power switch related to the design.
|
|