• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SET MENISCUS MENDER II DISPOSABLE; PASSER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SET MENISCUS MENDER II DISPOSABLE; PASSER Back to Search Results
Model Number 7209485
Device Problem Fracture (1260)
Patient Problem No Patient Involvement (2645)
Event Date 07/03/2020
Event Type  malfunction  
Event Description
It was reported that during meniscus repair when the package was opened, it was found that the needle and head were separated.It is unknown if there was a delay and a back up was available to complete the procedure.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
One 7209485 disposable meniscus mender ii set was returned for evaluation.Visual assessment confirmed the complaint.The loop meniscal were broken from the bottom.To avoid components from shifting within the package, individual component storage cradles are intentionally snug.Use of the distal end (head) to remove snare loops from the tray may result in weakening, bending or complete fracture between at the head and shaft connection.Proper method of retrieval is to push and pop the head of the component from the back of the cradle.Complaint history review indicated similar allegations for the lot number reported.Batch review did not indicate a condition, product or procedure failure that supported the allegation although the condition was confirmed during further investigation by engineering.This resulted with initiation of a corrective action and potential process change.Engineering evaluation confirmed the product met specifications at the time of distribution.Allegation rate of occurrences continue to be monitored via surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SET MENISCUS MENDER II DISPOSABLE
Type of Device
PASSER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key10418603
MDR Text Key203274602
Report Number1219602-2020-01260
Device Sequence Number1
Product Code HWQ
UDI-Device Identifier03596010468482
UDI-Public03596010468482
Combination Product (y/n)N
PMA/PMN Number
K885311
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7209485
Device Catalogue Number7209485
Device Lot Number2033453
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2020
Initial Date Manufacturer Received 07/23/2020
Initial Date FDA Received08/18/2020
Supplement Dates Manufacturer Received09/03/2020
Supplement Dates FDA Received09/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age27 YR
-
-