• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL BRK TRANSSEPTAL NEEDLE, 71 CM LENGTH; TROCAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL BRK TRANSSEPTAL NEEDLE, 71 CM LENGTH; TROCAR Back to Search Results
Model Number 407200
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Death (1802); Cardiac Perforation (2513)
Event Date 07/25/2020
Event Type  Death  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported tamponade could not be conclusively determined.Per the ifu, cardiac perforation is a known risk during the use of this device.Further information regarding the event was requested but not received.
 
Event Description
Related manufacturer reference: 3005334138-2020-00362, 3008452825-2020-00416.During a ventricular tachycardia ablation procedure, a cardiac tamponade occurred.Following the transseptal puncture, when attempting the retro aortic technique, a cardiac tamponade was observed.The tamponade was confirmed by echocardiogram located at the left atrial appendage.It is unknown which device caused the tamponade.Defibrillation, cardiac massage and cardiac surgery were performed to treat the patient.The patient expired due to instable hemodynamics.There were no performance issues with any abbott device.
 
Event Description
During a ventricular tachycardia ablation procedure, a cardiac tamponade occurred.A transseptal puncture was performed using guided intracardiac echocardiogram and a transesophageal echocardiogram due to anatomical changes in the interatrial septum (hypertrophy).Difficulty was noted maneuvering through the left atrium to the left ventricular with the flexability catheter and the md opted to use a retrograde approach.During this approach, a cardiac tamponade was diagnosed by echocardiogram requiring at thoracotomy for drainage.The patient experienced severe hemorrhagic shock, severe systemic dysfunction refractory to measures of blood products, vasoactive drugs and intra-aortic balloon pump resulting in death.It was unknown when the perforation occurred, or which device caused the tamponade.There were no performance issues with any abbott device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK TRANSSEPTAL NEEDLE, 71 CM LENGTH
Type of Device
TROCAR
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
MDR Report Key10430841
MDR Text Key203659235
Report Number3008452825-2020-00417
Device Sequence Number1
Product Code DRC
UDI-Device Identifier05414734205092
UDI-Public05414734205092
Combination Product (y/n)N
PMA/PMN Number
K072278
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Model Number407200
Device Catalogue Number407200
Device Lot Number7159278
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/28/2020
Initial Date FDA Received08/20/2020
Supplement Dates Manufacturer Received09/18/2020
Supplement Dates FDA Received09/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
FLEXABILITY ABLATION CATHETER, BI-D, CURVE F-J; SWARTZ BRAIDED TRANSSEPTAL INTRODUCER; FLEXABILITY ABLATION CATHETER, BI-D, CURVE F-J; SWARTZ BRAIDED TRANSSEPTAL INTRODUCER
Patient Outcome(s) Death;
-
-