• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND GMBH HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND GMBH HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Defective Device (2588)
Patient Problem No Patient Involvement (2645)
Event Date 07/24/2020
Event Type  malfunction  
Manufacturer Narrative
There was no known patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402.Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in (b)(6).Field engineer investigated the device and replaced the patient pump motor-1 and distribution pcb because both are faulty.Following testing has been performed and positive results allow to release the device in service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova received a report about a heater cooler alarming e17.No evidence of any patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key10435902
MDR Text Key207736905
Report Number9611109-2020-00477
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 07/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number16-02-80
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/24/2020
Initial Date FDA Received08/21/2020
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-