• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MI Z HANDLE ACET REAMER; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MI Z HANDLE ACET REAMER; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71364073
Device Problem Disconnection (1171)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/23/2020
Event Type  malfunction  
Event Description
It was reported that, during surgery, the device sheared off the reamer connection of the reamer shaft while reaming the acetabulum.The procedure was finished using a smith and nephew backup device.No delay or harm reported.
 
Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MI Z HANDLE ACET REAMER
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10436708
MDR Text Key203838211
Report Number1020279-2020-04121
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010507570
UDI-Public03596010507570
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71364073
Device Catalogue Number71364073
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 07/24/2020
Initial Date FDA Received08/21/2020
Supplement Dates Manufacturer Received03/22/2021
Supplement Dates FDA Received03/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-