• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. RSP; BASEPLATE, GLENOID HA-COAT, RSP, 6.5MM X 30MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. RSP; BASEPLATE, GLENOID HA-COAT, RSP, 6.5MM X 30MM Back to Search Results
Model Number 508-32-104
Device Problems Defective Component (2292); Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/24/2020
Event Type  Injury  
Manufacturer Narrative
The reason for this revision surgery due to baseplate failure.The previous surgery and the surgery detailed in this event occurred 4 years and 8 months apart.The healthcare professional indicated that there was significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements and showed no non-conforming material reports (ncmr) associated with the main contributor component listed in the complaint.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to baseplate failure.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical such as poor bone density, patient bone deterioration, inadequate soft tissue support, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event Description
Revision surgery - baseplate failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP
Type of Device
BASEPLATE, GLENOID HA-COAT, RSP, 6.5MM X 30MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445,
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445,
Manufacturer Contact
teffany hutto
9800 metric blvd
austin, tx 78758-5445, 
MDR Report Key10436814
MDR Text Key203868062
Report Number1644408-2020-00670
Device Sequence Number1
Product Code PHX
UDI-Device Identifier00888912024754
UDI-Public(01)00888912024754
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051075
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/30/2021
Device Model Number508-32-104
Device Catalogue Number508-32-104
Device Lot Number866C2019
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/24/2020
Initial Date FDA Received08/21/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/28/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
506-03-114 LOT 831C1261; 506-03-114 LOT 831C1262; 506-03-126 LOT 834C1217; 506-03-126 LOT 834C1217; 508-32-101 LOT 862C2057; 509-00-432 LOT 385P1011
Patient Outcome(s) Other; Required Intervention;
Patient Age68 YR
-
-