• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS PIIC IX HARDWARE; CENTRAL STATION MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS PIIC IX HARDWARE; CENTRAL STATION MONITOR Back to Search Results
Model Number 866424
Device Problems No Audible Alarm (1019); Disconnection (1171)
Patient Problems Death (1802); Fall (1848)
Event Date 08/14/2020
Event Type  Death  
Manufacturer Narrative
A follow up report will be submitted once the investigation is complete.
 
Event Description
The customer reported that they did not hear ecg or spo2 disconnect alarms at the central station for a patient that fell, and ultimately expired, around 17:00 on (b)(6) 2020.
 
Manufacturer Narrative
There was no product malfunction.This is considered a configuration issue, as the alarms did occur, but were set at a volume the user couldn't hear.The alarm volume for yellow and technical alarms was increased by the customer biomedical engineer to resolve the issue.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIIC IX HARDWARE
Type of Device
CENTRAL STATION MONITOR
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
MDR Report Key10440492
MDR Text Key203973007
Report Number1218950-2020-04868
Device Sequence Number1
Product Code MHX
UDI-Device Identifier00884838052130
UDI-Public(01)00884838052130
Combination Product (y/n)N
PMA/PMN Number
K102495
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866424
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/19/2020
Initial Date FDA Received08/22/2020
Supplement Dates Manufacturer Received08/19/2020
Supplement Dates FDA Received10/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age74 YR
-
-