The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction tubing leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot j311 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot j311 shows no trends.Trends were reviewed for complaint category, tubing leak.No trends were detected for this complaint category.The complaint kit with smart card was returned for investigation.A review of the data recorded on the smart card data showed that the treatment proceeded until an alarm #52: collect line air detected was received after 274 ml of whole blood had been processed.Visual inspection of the returned kit found a scuff on the return pump tubing segment.The tubing segment was pressure tested to check for leaks and a leak was verified at the site of the scuff.The damage to the tubing is at the location where there is an edge on the instrument pump head.The size, shape, and location of the observed damage on the pump loop is consistent with damage that occurs if the pump loop tubing is inadvertently rubbed against the pump head during installation by the end user.All cellex kits are leak tested prior to packaging.A leak of this nature would have been detected during in-process testing; therefore, it is unlikely the tubing damage was present at the time of manufacture.The root cause of the damage to the pump loop tubing was most likely caused during installation of the pump loop by the end user.No manufacturing related defects were confirmed during the evaluation.No further action is required at this time.(b)(4).
|