(b)(4).Visual analysis of the returned device found the leading tip and distal part of coil were completely stripped of coating and the coil was deformed.The tip of the device was missing and the exposed end appeared to be melted and discolored.Based on all available information, it is most likely that the user fired upon the coil with a laser causing it to detach and had to remove the coating of the coil in order to remove the device.The tip of the device was missing, including part of the coil and the distal tip of the device, and melting and discoloration was evident.The found damage was consistent with laser damage.The directions for use states "warning: to minimize risk of device breakage or patient injury, do not fire laser directly on any part of the stone cone coil.Warning: do not set holmium laser energy above.8 joules or power setting over 8 watts." therefore, the most probable root cause is unintended use error caused or contributed to event.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly, and performance specifications at the time of release for distribution.A labeling review was performed, and there is no evidence that the device was used not in accordance with the labeling.
|
It was reported to boston scientific corporation that a stone cone nitinol retrieval coil was used during a transurethral lithotomy procedure performed on (b)(6) 2020.According to the complainant, during the procedure, the coil was difficult to open and close.Additionally, it was reported that a part of the device was damaged.The procedure was completed with a different device.There were no patient complications reported as a result of this event.The investigation results revealed the coil/cone peeled/sheared and the tip separated; therefore, this is now an mdr reportable event.
|