Model Number VICMO12.6
Device Problems Off-Label Use (1494); No Apparent Adverse Event (3189)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/14/2020
Event Type  malfunction  
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.(b)(4).No similar complaint was reported for units within the same lot.(b)(4).
 
Event Description
The reporter indicated the surgeon inserted and removed a 12.6mm vicmo12.6 implantable collamer lens, -07.50 diopter, within the same surgery due to the lens tore during delivery into the patients left eye (os).There was no patient injury.The lens was exchanged for another same model/length lens within the same surgery and the problem was resolved.Cause of the event was unknown.
 
Manufacturer Narrative
Additional information: h3- device evaluation: lens was returned in a micro centrifuge vial with moisture and debris on the lens.Visual inspection found the optic and haptic torn with foreign debris on the lens.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia, ca
MDR Report Key10445865
MDR Text Key204196601
Report Number2023826-2020-01817
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 07/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2023
Device Model NumberVICMO12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2020
Initial Date Manufacturer Received 07/29/2020
Initial Date FDA Received08/24/2020
Supplement Dates Manufacturer Received12/05/2020
Supplement Dates FDA Received01/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CARTRIDGE MODEL SFC-45 - LOT # 1463923; FOAM TIP PLUNGER MODEL FTP - LOT # 1460510; INJECTOR MODEL MSI-PF - LOT # 1395302
Patient Age25 YR
-
-