Additional information was received that the medicinal solution that leaked was heparinized saline.The patient was a girl, 1 year, 8 months years old.One triple lumen pediasat catheter was returned for evaluation.A non-edwards 10ml syringe was attached at the distal lumen hub.Customer report of leakage issue was not able to be confirmed during evaluation due to returned sample condition.As received into lab, the catheter body was cut in half at the distal end of the backform.It was reported that the catheter was cut off by the customer.Leak test was performed to the sample from hubs to distal end of backform, and no leakage was observed.It was not able to perform leak test to the catheter body side because it was cut off.No visible damage was observed at the catheter tip bond area.No other visible damage was observed on the catheter body during visual examination.Visual examinations were performed under microscope at 10x magnification.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.If a leak in the backform is undetected at routine pre-insertion flushing, the catheter could leak during use and need to be removed/ exchanged for a new one.This could be done over a guidewire with minimal delay in therapy, and would not require a new venous access.These catheters, however, are used in pediatric patients, who have a lower threshold for brief interruptions in therapy or medication loss than adult patients.This results in a higher potential for an injury to occur.Therefore this is a reportable event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|