SMITH & NEPHEW, INC. MI Z HANDLE ACET REAMER; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
Model Number 71364073 |
Device Problem
Break (1069)
|
Patient Problems
Failure of Implant (1924); No Information (3190); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/03/2020 |
Event Type
Injury
|
Event Description
|
It was reported that during thr surgery while reaming, the mi z handle acet reamer gear shaft broke.When the back up reamer was used, the gears broke in that one as well.Surgeon was able to use a straight reamer to successfully complete the procedure with minimal time lost.Any other issue that could affect the patient's condition was not reported by this event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical / medical investigation concluded that, this complaint from the united states reports that the offset reamer gear shaft broke during a procedure.Surgeon was able to use a straight reamer to successfully complete the procedure with minimal time lost.No report of harm or impact to the patient.Smith and nephew has not received adequate materials (operative reports) to fully evaluate the complaint, but if additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.A complaint history review found related failures for the listed device; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|