Corrected data: see d.4., d.9., d10., & h.3.Manufacturer narrative: the reason for this complaint was reported as the screws on the baseplate stripped out.The healthcare professional indicated there was a significant adverse event to the patient.There was 5 minutes delay in surgery and another suitable device was available for use.The surgery was completed as intended.The instrument was inspected prior to surgery and was found to be acceptable.Examination confirmed the complaint, the returned instruments show the hexed/boxed ends are worn, corners rounded.The reported condition could possibly be a result of heavy use, misuse, wear and care must be taken to avoid compromising their performance, refer to the ifu.A review of the device history record (dhr) revealed, when released for use, the items met design and manufacturing requirements.Complaint database review showed previous complaints but there were no indications that these instruments have a design or material deficiency.The root cause of this complaint is attributable to damage incurred from prolonged use and through misuse or rough handling which surgical instruments are subjected to.This is not an event associated with a product failure, malfunction or issue.There are no indications that these instruments have a design or material deficiency therefore no containment of inventory is required.Event is associated with instrument usage, not a design or manufacturing issue.Surgical instruments condition can be determined while being used for its intended purpose.The replacement of surgical instruments due to normal wear and tear does not indicate a product deficiency, failure or issue.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
|