Model Number 3100 A
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Patient Involvement (2645)
Event Date 07/29/2020
Event Type  malfunction  
Manufacturer Narrative
Vyaire file identification: (b)(4).The reported issue was also by rt dept while doing the patient circuit calibration.As a resolution, the customer will replaced the panel meter.The suspect/device component has not been returned to vyaire.Therefore, a definitive root cause cannot be determined.Any additional information received from the customer will be included in a follow-up report.
 
Event Description
It is reported to vyaire medical that 3100a ventilator experienced a fluctuating map between 53-58 cm on the display while checking the zero/span on this unit.The external meter is reading stable at 37 cm.The customer confirmed that there was no patient involvement associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa, il
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs, ca
Manufacturer Contact
peter sparacio
26125 n. riverwoods blvd.
mettawa, il 
3273284
MDR Report Key10451808
MDR Text Key204354566
Report Number2021710-2020-12332
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003024
UDI-Public(01)10846446003024(11)20100203
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100 A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/29/2020
Initial Date FDA Received08/25/2020
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/03/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-