The patient was undergoing a thrombectomy procedure in the m1 segment of the middle cerebral artery (mca) using a penumbra engine canister (canister), aspiration tubing (tubing), a penumbra engine (engine), a penumbra system ace 60 reperfusion catheter (ace60), and a non-penumbra stent retriever.During the procedure, the tubing was connected to the patient and aspiration was initiated.However, none of the engine indicator lights were illuminated.The physician reset the engine but the issue was not resolved.Next, the physician decided to remove the canister and exchange it for a new one.It was also reported that the lid of the canister felt loose.The procedure was then completed using the new canister, the same engine, the same tubing, the same ace60, and the same stent retriever.There was no report of an adverse effect to the patient.
|
Results: the canister was undamaged.Conclusions: evaluation of the returned canister could not confirm the reported complaint.During functional testing, the canister was seated onto a demonstration engine and the engine was able to achieve vacuum pressure within specification and all four vacuum indicator lights illuminated.Penumbra pump canisters are 100% visually inspected and functionally tested during in-process inspection by the supplier.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|