• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH SIZE 8 ACCOLADE II 132 DEG; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH SIZE 8 ACCOLADE II 132 DEG; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 6720-0837
Device Problem Unstable (1667)
Patient Problems Injury (2348); Joint Laxity (4526)
Event Date 02/13/2020
Event Type  Injury  
Manufacturer Narrative
Review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event Description
Procedure: the patient underwent right total hip arthroplasty (b)(6) 2020.Patient underwent revision right total hip for instability on (b)(6) 2020.Head, liner and stem exchanged.
 
Manufacturer Narrative
Reported event: an event regarding revision involving an accolade stem was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned.Clinician review: a review of the provided medical records by a clinical consultant stated that : one month after index tha the patient underwent revision surgery for instability.It is unclear why the stem needed to be exchanged.On the post revision xray there is a circlage wire, it is not known if this was for prophylaxis against, or treatment for , calcar fracture.Further clarification from the clinician noted," dislocation of a right pressfit tha is confirmed by pre-operative x-rays.Without further confirmatory information such as an operative report and implant sheet it is not possible to be certain that the stem was revised.The stem has the same appearance and is in the same position as the pre-operative stem.However, post revision x-rays show the presence of a cerclage wire around the proximal femur.This wire would have been placed for one of two reasons.1.For prophylaxis against fracture when attempting to remove the stem or when inserting a new stem.2.A fracture did occur in the process of removal or insertion of a stem.That being stated without the documentation noted above revision of the stem can not be confirmed." device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: it was reported that the patient was revised due to instability.A review of the provided medical records by a clinical consultant stated that : one month after index tha the patient underwent revision surgery for instability.It is unclear why the stem needed to be exchanged.On the post revision xray there is a circlage wire, it is not known if this was for prophylaxis against, or treatment for , calcar fracture.Further clarification from the clinician noted that dislocation of a right pressfit tha is confirmed by pre-operative x-rays.Without further confirmatory information such as an operative report and implant sheet it is not possible to be certain that the stem was revised.The stem has the same appearance and is in the same position as the pre-operative stem.However, post revision x-rays show the presence of a cerclage wire around the proximal femur.This wire would have been placed for one of two reasons.1.For prophylaxis against fracture when attempting to remove the stem or when inserting a new stem.2.A fracture did occur in the process of removal or insertion of a stem.That being stated without the documentation noted above revision of the stem can not be confirmed.The event could not be confirmed nor the exact cause be determined because insufficient information was provided.Additional information including operative reports, progress notes and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.H3 other text : device not returned.
 
Event Description
Procedure: the patient underwent right total hip arthroplasty (b)(6) 2020.Patient underwent revision right total hip for instability on (b)(6) 2020.Head, liner and stem exchanged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIZE 8 ACCOLADE II 132 DEG
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key10458126
MDR Text Key204538496
Report Number0002249697-2020-01745
Device Sequence Number1
Product Code LPH
UDI-Device Identifier04546540664518
UDI-Public04546540664518
Combination Product (y/n)N
PMA/PMN Number
K143085
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 01/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number6720-0837
Device Catalogue Number6720-0837
Device Lot Number60145302
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/03/2020
Initial Date FDA Received08/27/2020
Supplement Dates Manufacturer Received12/22/2020
Supplement Dates FDA Received01/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age48 YR
Patient Weight95
-
-