• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OBERDORF SYNTHES PRODUKTIONS GMBH 2.0MM IMF SCREW SELF-DRILLING 12MM; SCREW FIXATION INTRAOSSEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OBERDORF SYNTHES PRODUKTIONS GMBH 2.0MM IMF SCREW SELF-DRILLING 12MM; SCREW FIXATION INTRAOSSEOUS Back to Search Results
Catalog Number 201.932
Device Problems Break (1069); Entrapment of Device (1212)
Patient Problem No Code Available (3191)
Event Date 07/25/2020
Event Type  Injury  
Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.Part number: 401.882.Lot number: 21p1995.Manufacturing site: (b)(4).Release to warehouse date: october 22, 2019.A manufacturing record evaluation was performed for the not sterile device lot number, and no non-conformance were identified.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2020, when placing the 12 mm imf screw, the head of the screw broke and a fragment remains in the patient¿s right lower jaw.There was a surgical delay of twenty (20) minutes.The procedure was completed successfully.The patient outcome was satisfactory.Concomitant devices reported: screwdriver (part number unknown, lot unknown, quantity 1).This report involves one (1) 2.0mm ti locking scr slf-tpng with stardrive recess 12mm.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: device history lot part number: 201.932, lot number: 21p1715, manufacturing site: mezzovico, release to warehouse date: 21 oct 2019.A manufacturing record evaluation was performed for the not sterile device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
This report involves one (1) 2.0mm imf screw self-drilling 12mm concomitant devices reported: unknown- screwdrivers(part# unknown, lot# unknown, quantity# 1).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.0MM IMF SCREW SELF-DRILLING 12MM
Type of Device
SCREW FIXATION INTRAOSSEOUS
Manufacturer (Section D)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key10459758
MDR Text Key204696168
Report Number8030965-2020-06371
Device Sequence Number1
Product Code DZL
UDI-Device Identifier07611819140899
UDI-Public(01)07611819140899
Combination Product (y/n)N
PMA/PMN Number
K010527
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 08/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number201.932
Device Lot Number21P1715
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/03/2020
Initial Date FDA Received08/27/2020
Supplement Dates Manufacturer Received08/27/2020
10/08/2020
10/13/2020
10/09/2020
Supplement Dates FDA Received09/22/2020
10/09/2020
10/13/2020
11/04/2020
Patient Sequence Number1
Treatment
IMF SCREW Ø2 L12 SST; UNK - SCREWDRIVERS; UNK - SCREWDRIVERS
Patient Outcome(s) Required Intervention;
Patient Age26 YR
-
-