ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number SN6AT5
Device Problem Unintended Movement (3026)
Patient Problem Visual Impairment (2138)
Event Date 07/17/2020
Event Type  Injury  
Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.Additional information has been requested.(b)(4).
 
Event Description
A physician reported a patient with poor vision of the left eye four days following toric intraocular lens (iol) implant.The lens was noted to have rotated from 76 degrees to 45 degrees.Nine days following the initial procedure the surgeon rotated the lens and a week later it rotated again.A capsular tension ring was placed in the left eye and the lens rotation was adjusted.Postoperatively the lens continued to rotate and an explant procedure is planned.Additional information is requested.There are two related reports for this patient.This report addresses the patient's left eye, and another manufacturer report will be filed for the fellow eye.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ TORIC SINGLEPIECE IOL
Type of Device
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514979
MDR Report Key10460903
MDR Text Key204680388
Report Number1119421-2020-01196
Device Sequence Number1
Product Code MJP
Combination Product (y/n)N
Reporter Country CodeAR
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 08/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2020
Device Model NumberSN6AT5
Device Catalogue NumberSN6AT5.085
Device Lot Number12402993
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/07/2020
Initial Date FDA Received08/27/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/22/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-