• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 801763
Device Problem Battery Problem (2885)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Per the manufacturer's technical support specialist, the issue occurred about a month before it was reported and the unit was being used normally.The batteries were replaced by a third party service but they were not holding the charge.The field service representative (fsr) observed the battery state indication light stayed red and would not turn to green even after it charged overnight.The fsr replaced the power manager printed circuit board assembly (pcba) and the batteries.The battery status light then turned green.The unit operated to the manufacturer's specifications.
 
Event Description
It was reported that the batteries on the heart lung machine (hlm) were not holding a charge.No other details regarding the nature of this event were provided.
 
Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed that the power manager printed circuit board (pcb) was not functional.The power manager board was installed on the power manager test fixture which is connected to a lab use only (luo) central control monitor (ccm).After completing the boot sequence, a 'system computer needs service' message appeared on the screen.This was observed with both luo ccm versions, all power sources connected, alternating current (ac) power only and battery only.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS- ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key10462702
MDR Text Key206793643
Report Number1828100-2020-00327
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier00886799000656
UDI-Public(01)00886799000656(11)181023
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 03/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801763
Device Catalogue Number801763
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/03/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/04/2020
Initial Date FDA Received08/27/2020
Supplement Dates Manufacturer Received09/03/2020
03/15/2021
Supplement Dates FDA Received09/24/2020
03/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-