There was no patient involvement.Livanova (b)(4) manufactures the s5 system.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.When pump was turned on after period of time, he found out that the stop link symbol appeared on control display and then went away and "no monitoring/control pump" messages appeared.The technician replaced the processor board of one pump and moved connector on the e/p pack.The issue was solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|