Results of investigation: the device, intended for use in treatment, was returned for evaluation.A visual inspection confirms the jaws have fractured off the device.The fractured piece was not returned with the device.The t-handle fractured in the jaws of the handle, likely from static bending or torsional overload.An overload fracture can occur if the mechanical loads applied to the device exceed the strength of the mater.The device was manufactured in 2012.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.
|