SMITH & NEPHEW, INC. CONTROL UNIT DYONICS 25; ARTHROSCOPE Back to Search Results
Model Number 7211010
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Patient Involvement (2645)
Event Date 08/06/2020
Event Type  malfunction  
Event Description
It was reported that, the "dyonics 25 control unit" had issues while inserting a new cassette.The unit displayed error message "cassette insertion/pressure error".After this, when tried to solve the issue, it seemed the sensor of insertion was damaged.No case reported; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, intended for use in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was a relationship found between the returned device and the reported incident.An analysis of the customer provided images found that a cassette insertion error was displayed onscreen and a exposed pressure port from the device.The complaint was confirmed and the root cause was associated with mechanical component failure.Factors, unrelated to the manufacturing and design of the device which could have contributed to the reported event, include pressure transducer malfunction or the latch keeper wear.A review of the device history records could not be completed since the provided product identification information did not match any known release of this part.A complaint history review found concluded this was a repeat issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTROL UNIT DYONICS 25
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10470018
MDR Text Key204900596
Report Number1643264-2020-01010
Device Sequence Number1
Product Code HRX
UDI-Device Identifier03596010543554
UDI-Public03596010543554
Combination Product (y/n)N
PMA/PMN Number
K171794
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup
Report Date 09/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7211010
Device Catalogue Number7211010
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/07/2020
Initial Date FDA Received08/31/2020
Supplement Dates Manufacturer Received09/25/2020
Supplement Dates FDA Received09/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-