Catalog Number 04625374160
Device Problem Incorrect Measurement (1383)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/10/2020
Event Type  malfunction  
Manufacturer Narrative
The reporter's product was requested for investigation and replacement product was sent to the reporter.The product has not been received at this time.If the product is returned in the future, a follow up report will be submitted.Routine retention testing is performed.Retention testing data is reviewed and appropriate actions are taken as needed.The reporter's meter was provided for investigation where it was tested using retention strips and retention controls.Testing results (qc range = 4.1 - 6.8 inr): qc 1: 5.3 inr, qc 2: 5.2 inr, qc 3: 5.3 inr.The obtained qc values were in the allowed range of the used combination strip lot - qc lot.All measurements were without error messages.The maximum difference between measurements was 2%.Occupation: the occupation is lay user/patient.
 
Event Description
The initial reporter stated he received discrepant results when testing with coaguchek xs meter serial number (b)(4).At 8:52 a.M., a sample from the patient was tested using the meter, resulting with a value of > 8.0 inr.At 10:20 a.M., a sample from the patient was tested using the meter, resulting with a value of 5.6 inr.At 11:22 a.M., a sample from the patient was tested using the meter, resulting with a value of 2.9 inr.The patient's therapeutic range is 2.0 - 3.0 inr.The patient's testing frequency is weekly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS SYSTEM
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key10471568
MDR Text Key204940679
Report Number1823260-2020-02138
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 08/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/30/2020
Device Catalogue Number04625374160
Device Lot Number42401022
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/10/2020
Initial Date FDA Received08/31/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
ASPIRIN; COUMADIN; FENOFIBRATE; FLOMAX; LASIX; MAGNESIUM OXIDE; POTASSIUM CHLORIDE; SYNTHROID; ZOCOR
Patient Age75 YR
-
-