• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP ITALIA S.R.L. - CRM FACILITY PARADYM; DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP ITALIA S.R.L. - CRM FACILITY PARADYM; DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER Back to Search Results
Model Number PARADYM DR 8550
Device Problems High impedance (1291); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Low impedance (2285)
Patient Problem Electric Shock (2554)
Event Date 08/13/2020
Event Type  Injury  
Event Description
Reportedly, technical services were called on (b)(6) 2020 because the ventricular lead impedance and the rv coil continuity of the subject icd were saturated at 3000 ohms.Non-physiological signals, leading to a shock delivery, were observed on the ventricular channel.Non-physiological signals on the atrial channel, leading to oversensing, were noticed when reviewing the episodes recorded in the device memory.The shock impedance was high (269 ohms) and the shock could not be completely delivered because the maximum shock duration was met.In addition, intermittent low atrial lead impedance measurements were recorded in the device memory.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADYM
Type of Device
DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER
Manufacturer (Section D)
SORIN GROUP ITALIA S.R.L. - CRM FACILITY
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR  92140
Manufacturer (Section G)
SORIN GROUP ITALIA SRL VIA CRESCENTINO 13040 SALUGGIA ITALY
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR   92140
Manufacturer Contact
laura ouaki
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR   92140
146013429
MDR Report Key10475562
MDR Text Key205081061
Report Number1000165971-2020-00577
Device Sequence Number1
Product Code MRM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 09/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/28/2012
Device Model NumberPARADYM DR 8550
Device Catalogue NumberPARADYM DR 8550
Device Lot Number2580
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date08/13/2020
Initial Date Manufacturer Received 08/13/2020
Initial Date FDA Received09/01/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/28/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-