Device analysis by mfr: the advancer, handshake connections, sheath, coil, burr and annulus were microscopically and visually examined.The device was returned intact, there was no separation on the device.Microscopic examination of the device revealed that the annulus was damaged and not rounded which is consistent with damage seen with the use of a rotawire.The coil has become stretched due to manipulation during the procedure.Inspection of the remainder of the device presented no other damage or irregularities.
|